pH
Range-2.000 to 19.000 pH or 0 to 14 in aqueous solutions
Resolution0.1/0.01/0.001 pH
Relative Accuracy±0.002 pH
Smart Stability™ and Smart Averaging™Resolution dependent / automatic
Auto-Cal™ and/or Manual Calibration1 to 5 points; US/NIST, DIN and custom buffers recognised
ISE
Range0 to 19900
Resolution1, 2 or 3 significant digits
Relative Accuracy± 0.2mV or 0.05%, whichever is greater
Smart Stability™ and Smart Averaging™Resolution dependent / automatic
Units M, mg/L, %, ppb or no units
Number of ISE Cal Points 1 to 5
mV
Range ±1999.9mV
ORP EH autocalibration to standard hydrogen electrode
Resolution 0.1
Relative Accuracy ±0.2mV or 0.05%, whichever is greater
Temperature
Range -5 to 105°C
Resolution 0.1 up to 99.9°C, 1.0 over 99.9°C
Relative Accuracy ±0.1°C
Temperature Compensation Auto/Manual
Display
Custom backlit LCD for easy viewing
Inputs/Outputs
Sensor Inputs pH/ISE – BNC
ATC – miniDIN-Locking
Reference – Pin tip
Stirrer – phono-jack
Special GLP Software Features
Number of Data Logging Points 200 points by time, stability or manual read with time/date stamp
Calibration Log Last 10 calibrations
User Method Storage 10 saved
Advanced Self-Test and Diagnostics Yes
GLP Electrode Verification Yes
Password Protection Yes
Internet Software Updates / Downloadable Methods Yes
Power
Universal power adaptor and battery power (4 x AA batteries)
Regulatory and Safety
CE, CSA, UL, TÜV, FCC class A limits
Environmental Operating Conditions
Ambient Operating Temperature 5 to 45°C
Relative Humidity 5 to 85% non-condensing
IP Rating IP54, splashproof and dustproof